Measuring Area X
900
Measuring Area Y
1600
Measuring Area Z
800

Pre-owned measuring devices can offer excellent performance when checked in accordance with strict criteria and refurbished. And that is exactly what we guarantee you with our ZEISS Originals systems. ZEISS Originals systems enable you to harness the full potential of your quality assurance and achieve reliable measuring results while sticking to your budget. If you are interested in seeing what's available, please don't hesitate to contact us.

Product note

CMM ACCURA II 9/16/8

construction year: 2018 refurbished: 2023

Coordinate measuring machines with the ZEISS Originals seal are refurbished used devices with components that may be made of granite.

Granite is a natural product and shows signs of wear (scratches, etc.). These signs of use do not constitute a defect that justifies a complaint. The promised measuring accuracy of the coordinate measuring machine is not affected by the traces of use.

CMM ACCURA 9/16/8 ZEISS CNC Coordinate Measuring Machine (CMM) in bridge design

  • measuring volume X = 900 mm, Y = 1600 mm, Z = 800 mm

  • Table Height TH = 752 mm

  • incl. DotScan preparation

  • glass-ceramic scales in all 3 axes

  • integrated passive vibration damping Specifications here below hold for

  • S1: active sensors VAST XT gold, VAST gold, VAST XTR gold

  • S2: passive sensor VAST XXT with RDS

  • temperature range 20 - 26 °C, remaining constant for 48 hours

  • maximum temperature gradients 1.0 K/h 2.0 K/d 1.0 K/m Max. Permiss. length measurement Error (MPE) as per ISO 10360-2:2009

  • MPE_E0 (3D) = (1.2 + L/350) µm for S1

  • MPE_E0 (3D) = (1.6 + L/350) µm for S2 Max. Permiss. Error (MPE) of scanning probing as per ISO 10360-4:2000 within Maximum Permissible Time (MPT) for scanning test

  • MPE_THP = 2.0 µm in MPT_tau = 40 s for S1

  • MPE_THP = 2.5 µm in MPT_tau = 50 s for S2 Single stylus form error as per ISO 10360-5:2010

  • MPE_PFTU = 1.2 µm for S1

  • MPE_PFTU = 1.7 µm for S2 Form measuring tolerance as per ISO 12181:2011 (VDI/VDE 2617 part 2.2)

  • MPE_RONt (MZCI) = 1.4 µm for S1

  • MPE_RONt (MZCI) = 1.7 µm for S2

  • pre-wired for tactile active and tactile passive sensors, for Multi Application Sensor System (mass)

  • VAST navigator and/or VAST XXT measurement technologies (tactile active scanning and/or tactile passive scanning)